Balloon angioplasty of the coarctation of the aorta can result in intimal or medial or even adventitial dissections as demonstrated by in vitro studies and animal experiments. As a typical sign of stretching of the aortic wall, patients complain of pain during the angioplasty procedure. In the literature, aortic wall rupture and ventricular fibrillation during the procedure are reported. Additional sudden deaths can occur within 40 hours after the procedure. Mortality ranges from 0.1% to 2.5%. By transesophageal echocardiography, monitoring of balloon angioplasty, control of the positioning of the balloon, and control of the results and detection of complications are possible. Intimal as well as medial dissections can be detected with observed healing for intimal but also medial dissections. In order to avoid the patient's discomfort, intraaortic ultrasound will be used in the future, when major methodological improvements are done. Computed tomography demonstrates medial dissections but is not able to visualize intimal dissections. Using computed tomography and magnetic resonance after angioplasty of the coarctation of the aorta, irregularities are described in up to 17% of the patients. For angiography, a low detection rate of medial dissections has to be expected, when not biplane angiographies of the whole thoracic aorta are performed. Medial dissections can be seen, but intimal dissections are missed. In conclusion, a review of the literature demonstrates a high incidence of intimal and medial dissections after angioplasty of the coarctation of the aorta with spontaneous healing in most patients. As is the way with coronary angioplasty, aortic wall ruptures are rare, but stand-by surgery is necessary.