Three-dimensional CT-guided bronchoscopy with a real-time electromagnetic position sensor: a comparison of two image registration methods

Chest. 2000 Dec;118(6):1783-7. doi: 10.1378/chest.118.6.1783.

Abstract

Study objectives: To compare two different image registration methods for accurately displaying the position of a flexible bronchoscope on a previously acquired three-dimensional CT scan during bronchoscopy.

Setting: Bronchoscopy suite of a university hospital.

Patients: Fifteen adult patients scheduled for nonemergent bronchoscopy.

Methods: A miniature electromagnetic position sensor was placed at the tip of a flexible bronchoscope. Previously acquired three-dimensional CT scans were registered with the patient in the bronchoscopy suite. Registration method 1 used multiple skin fiducial markers. Registration method 2 used the inner surface of the trachea itself for registration. Method 1 was objectively assessed by measuring the error distance between the real skin marker position and the computer display position. Methods 1 and 2 were subjectively assessed by the bronchoscopist correlating visual bronchoscopic anatomic location with the computer display position on the CT image.

Results: The error distance (+/- SD) from known points for registration method 1 was 5.6 +/- 2.7 mm. Objective error distances were not measured for method 2 because no accurate placement of the bronchoscope sensor could be correlated with CT position. Subjectively, method 2 was judged more accurate than method 1 when compared with the fiberoptic view of the airways through the bronchoscope. Additionally, method 2 had the advantage of not requiring placement of fiducial markers before the CT scan. Respiratory motion contributed an error of 3.6 +/- 2.6 mm, which was partially compensated for by a second tracking sensor placed on the patient's chest.

Conclusion: Image registration method 2 of surface fitting the trachea rather than method 1 of fiducial markers was subjectively judged to be superior for registering the position of a flexible bronchoscope during bronchoscopy. Method 2 was also more practical inasmuch as no special CT scanning technique was required before bronchoscopy.

Publication types

  • Comparative Study

MeSH terms

  • Biopsy*
  • Bronchoscopy*
  • Electromagnetic Phenomena*
  • Female
  • Humans
  • Male
  • Radiography, Interventional*
  • Tomography, X-Ray Computed*