Blinded, multicenter quality control study for the quantification of human immunodeficiency virus type 1 RNA in plasma by the Belgian AIDS reference laboratories

Clin Microbiol Infect. 2000 Apr;6(4):213-7. doi: 10.1046/j.1469-0691.2000.00048.x.

Abstract

Objective: In order to evaluate the interlaboratory variation of HIV-1 RNA measurements in plasma, the Belgian AIDS reference laboratories organized a blinded multicenter quality control study.

Methods: Atest panel of coded spiked HIV-1 plasma samples reflecting the dynamic range of the assay was composed and distributed. The HIV-1 RNA concentration of these samples was determined by the eight Belgian AIDS reference laboratories by means of the Amplicor HIV-1 Monitor version 1.5 assay.

Results: Analysis of the results demonstrated that there was little interlaboratory variation for the high concentration range (4.0-5.7 log10 copies/mL), never exceeding 0.2 log10 copies/mL. However the standard deviation for the low concentration range (2.6-3.9 log10 copies/mL) reached up to 0.22 log10 copies/mL.

Conclusions: Since interlaboratory variability never reached 0.5 log10 copies/mL and each of the laboratories was able to detect four-fold differences in plasma HIV-1 RNA levels, the Amplicor assay can be used in multicenter studies without a centralized analysis of samples. Furthermore, this well-characterized proficiency panel of spiked plasma samples could be used as a standard in the study of interassay comparisons.

Publication types

  • Evaluation Study
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Belgium
  • HIV Infections / blood
  • HIV Infections / virology*
  • HIV-1 / genetics
  • HIV-1 / isolation & purification*
  • Humans
  • Laboratories / standards*
  • Polymerase Chain Reaction / standards*
  • Quality Control
  • RNA, Viral / blood*
  • Reference Values
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Viral Load