Objective: In order to evaluate the interlaboratory variation of HIV-1 RNA measurements in plasma, the Belgian AIDS reference laboratories organized a blinded multicenter quality control study.
Methods: Atest panel of coded spiked HIV-1 plasma samples reflecting the dynamic range of the assay was composed and distributed. The HIV-1 RNA concentration of these samples was determined by the eight Belgian AIDS reference laboratories by means of the Amplicor HIV-1 Monitor version 1.5 assay.
Results: Analysis of the results demonstrated that there was little interlaboratory variation for the high concentration range (4.0-5.7 log10 copies/mL), never exceeding 0.2 log10 copies/mL. However the standard deviation for the low concentration range (2.6-3.9 log10 copies/mL) reached up to 0.22 log10 copies/mL.
Conclusions: Since interlaboratory variability never reached 0.5 log10 copies/mL and each of the laboratories was able to detect four-fold differences in plasma HIV-1 RNA levels, the Amplicor assay can be used in multicenter studies without a centralized analysis of samples. Furthermore, this well-characterized proficiency panel of spiked plasma samples could be used as a standard in the study of interassay comparisons.