Background: The aortic Carpentier-Edwards pericardial bioprosthesis offers good long-term clinical outcomes with a low rate of structural deterioration. However, little in vivo hemodynamic data is available for this bioprosthesis.
Methods: To determine the hemodynamic performance of the 19-mm Carpentier-Edwards pericardial valve, both cardiac catheterization and dobutamine stress echocardiography were electively performed in 10 patients. The mean age at the study was 71.6 +/- 4.4 years and the mean body surface area was 1.39 +/- 0.11 m2. The peak-to-peak gradient, instantaneous peak gradient, mean gradient, and valve orifice area were measured by standard cardiac catheterization. The Doppler-derived gradients and valve orifice area were also measured both at rest and during dobutamine infusion.
Results: The average peak-to-peak gradient, instantaneous peak gradient, mean gradient, and valve orifice area measured by catheterization were 13.0 +/- 5.4 mmHg, 28.5 +/- 7.7 mmHg, 12.0 +/- 4.9 mmHg, and 1.55 +/- 0.45 cm2, respectively. The peak and mean Doppler gradients, and valve orifice area by resting echocardiography were 27.7 +/- 9.5 mmHg, 12.3 +/- 4.8 mmHg, and 1.39 +/- 0.26 cm2, respectively. At a dosage of 10 microg/kg/min of dobutamine, the mean Doppler gradient rose mildly to 22.2 +/- 4.8 mmHg, while the cardiac output increased from 4.49 +/- 0.44 to 6.64 +/- 0.87 L/min. The valve orifice area during the 10 microg/kg/min dobutamine infusion (1.55 +/- 0.25 cm2) was significantly larger than its value at rest (p < 0.05).
Conclusions: With acceptable hemodynamic performance, use of the aortic 19-mm Carpentier-Edwards pericardial valve is a reliable option for elderly patients with a small annulus.