A practical approach to inverse planning for high-precision dose escalated conformal prostate radiotherapy

Phys Med Biol. 2001 May;46(5):1473-85. doi: 10.1088/0031-9155/46/5/310.

Abstract

The problem of choosing the best gantry angles and beam weights for dose-escalated conformal prostate treatment planning is formulated using a mixed-integer linear programming approach, to account for tumour dose homogeneity and dose-volume constraints. The formulation allows the number of beams to be restricted and for some of the beams to be compulsory. The present planning algorithm interfaces with and utilizes the three-dimensional planning capabilities of a commercial treatment planning system. A case study is illustrated, which represents a particularly challenging planning problem due to a large planning target volume and an unusually small bladder. Treatment plans with different numbers of beams are generated to compare with each other and with the standard six-field plan. Significant improvement is shown in the reduction of hot regions within the femoral heads and rectal wall, while not unduly compromising homogeneity constraints for the tumour.

Publication types

  • Comparative Study

MeSH terms

  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Prostatic Neoplasms / radiotherapy*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted*
  • Radiotherapy, Conformal / methods*
  • Software
  • Urinary Bladder / anatomy & histology