RF inhomogeneity compensation in structural brain imaging

Magn Reson Med. 2002 Feb;47(2):398-402. doi: 10.1002/mrm.10050.

Abstract

Three-dimensional T(1)-weighted magnetization-prepared rapid gradient-echo (MP-RAGE) sequences with centric phase encoding (PE) in the inner loop provide structural brain images with a high spatial resolution and high tissue contrast. A disadvantage of this sequence type is the susceptibility to inhomogeneities of the radiofrequency (RF) coil, which may result in poor image contrast in some peripheral regions. A special excitation pulse is presented which compensates for these effects in both the head/foot and anterior/posterior directions. This pulse has a duration of only 1.3 ms and is thus compatible with the short repetition times (TRs) required for MP-RAGE imaging. It is shown experimentally that images acquired with the compensation pulse may be segmented without using intensity correction algorithms during data postprocessing.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts
  • Brain / anatomy & histology*
  • Echo-Planar Imaging / methods*
  • Humans
  • Image Enhancement / methods*
  • Image Processing, Computer-Assisted*
  • Reference Values
  • Sensitivity and Specificity