Direct stent implantation using the EXPRESS Coronary Stent System: results of a multi-center feasibility study

J Interv Cardiol. 2003 Dec;16(6):491-7. doi: 10.1046/j.1540-8183.2003.01057.x.

Abstract

The aim of this prospective, multicenter, single arm study was to assess the safety and feasibility of EXPRESS Coronary Stent implantation in native coronary arteries without balloon predilatation. Forty-two patients with de novo or restenotic lesions were enrolled, of which 38 were eligible for analysis. The coronary lesions were predominantly complex, occurring in arteries with a mean reference diameter of 2.80 +/- 0.49 mm. Technical and procedural success were achieved in 89.5% and 84% of the cases respectively. The mean minimal lumen diameter increased from 1.08 +/- 0.26 mm to 2.55 +/- 0.44 mm and diameter stenosis decreased from 61 +/- 7% to 13 +/- 8%. The primary endpoint of major adverse cardiac events at 30 days was 2.6% and was limited to only one event (target vessel revascularization, nontarget lesion). No other MACE were observed during the three-month follow-up period. Based on the findings of this study, direct stenting with the EXPRESS Stent appears feasible and is well tolerated.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary
  • Coronary Angiography
  • Coronary Restenosis / diagnostic imaging
  • Coronary Restenosis / etiology
  • Coronary Restenosis / prevention & control*
  • Coronary Stenosis / complications
  • Coronary Stenosis / diagnostic imaging
  • Coronary Stenosis / therapy*
  • Equipment Design
  • Feasibility Studies
  • Female
  • Follow-Up Studies
  • Humans
  • Male
  • Myocardial Infarction / diagnostic imaging
  • Myocardial Infarction / etiology
  • Myocardial Infarction / prevention & control*
  • Myocardial Revascularization
  • Prospective Studies
  • Stents*
  • Treatment Outcome