Abstract

We sought to determine the hemodynamic performance of the Carpentier-Edwards Perimount pericardial valve in the mitral position. We reviewed the Doppler echocardiographic data on 189 patients (110 women; 68 +/- 12 years of age) who were implanted with this valve (7.6 days +/- 13 postoperatively) at our institution between September 2000 and May 2002. The average ejection fraction was 47%. For all valves, the peak velocity was 1.9 +/- 0.3 m/s, peak gradient was 15 +/- 4.8 mm Hg, and mean gradient was 5.8 +/- 2 mm Hg. The pressure half-time was 93 +/- 24 milliseconds, with a calculated effective orifice area of 2.5 +/- 0.6 cm2. The average effective orifice area by continuity equation (83 valves) was 1.5 +/- 0.5 cm2. The mitral regurgitation was graded mild or less in 97.5% of all valves. This is the largest series establishing the favorable hemodynamic behavior of the different sizes of a new Perimount mitral valve, and the reported data could serve as a reference.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Bioprosthesis*
  • Coronary Circulation / physiology*
  • Echocardiography, Doppler / methods*
  • Female
  • Heart Valve Prosthesis*
  • Hemodynamics / physiology
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / diagnostic imaging*
  • Time Factors