Accuracy of MR temperature measurement based on chemical shift change for radiofrequency ablation using hook-shaped electrodes

Magn Reson Med Sci. 2004 Jul 15;3(2):95-100. doi: 10.2463/mrms.3.95.

Abstract

The purpose of the current study is to evaluate the accuracy of MR thermometry for radiofrequency ablation (RFA) with hook-shaped electrodes. The objects were eight extracted bovine livers. The chemical shift change was calculated from MR images acquired with a spoiled gradient echo sequence and compared with the temperature directly measured with a thermocouple. Linear regression was established between them with a coefficient of -0.0110+/-0.0007 ppm/ degrees C and errors were calculated as -0.50+/-7.50 degrees C. MR thermometry is capable of monitoring temperature for RFA.