Lung volume assessment for a cross-comparison of two breathing-adapted techniques in radiotherapy

Int J Radiat Oncol Biol Phys. 2005 Oct 1;63(2):602-9. doi: 10.1016/j.ijrobp.2005.05.020.

Abstract

Purpose: To assess the validity of gated radiotherapy of lung by using a cross-check methodology based on four-dimensional (4D)-computed tomography (CT) exams. Variations of volume of a breathing phantom was used as an indicator.

Methods and materials: A balloon was periodically inflated and deflated by a medical ventilator. The volume variation (DeltaV) of the balloon was measured simultaneously by a spirometer, taken as reference, and by contouring 4D-CT series (10 phases) acquired by the real-time position management system (RPM). Similar cross-comparison was performed for 2 lung patients, 1 with free breathing (FB), the other with deep-inspiration breath-hold (DIBH) technique.

Results: During FB, DeltaV measured by the spirometer and from 4D-CT were in good agreement: the mean differences for all phases were 8.1 mL for the balloon and 10.5 mL for a patient-test. End-inspiration lung volume has been shown to be slightly underestimated by the 4D-CT. The discrepancy for DeltaV between DIBH and end-expiration, measured from CT and from spirometer, respectively, was less than 3%.

Conclusions: Provided that each slice series is correctly associated with the proper breathing phase, 4D-CT allows an accurate assessment of lung volume during the whole breathing cycle (DeltaV error <3% compared with the spirometer signal). Taking the lung volume variation into account is a central issue in the evaluation and control of the toxicity for lung radiation treatments.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Humans
  • Lung / diagnostic imaging
  • Lung / physiopathology
  • Lung Neoplasms / radiotherapy*
  • Movement*
  • Phantoms, Imaging
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Respiration*
  • Spirometry / methods*
  • Tomography, X-Ray Computed / methods*