A hypofractionation treatment protocol for prostate cancer was initiated in our department in December 2003. The treatment regimen consists of a total dose of 36.25 Gy delivered at 7.25 Gy per fraction over 10 days. We discuss the rationale for such a prostate hypofractionation protocol and the need for frequent prostate imaging during treatment. The CyberKnife (Accuray Inc., Sunnyvale, CA), a linear accelerator mounted on a robotic arm, is currently being used as the radiation delivery device for this protocol, due to its incorporation of near real-time kV imaging of the prostate via 3 gold fiducial seeds. Recently introduced conventional linac kV imaging with intensity modulated planning and delivery may add a new option for these hypofractionated treatments. The purpose of this work is to investigate the use of intensity modulated radiotherapy (IMRT) and the Varian Trilogy Accelerator with on-board kV imaging (Varian Medical Systems Inc., Palo Alto, CA) for treatment of our hypofractionated prostate patients. The dose-volume histograms and dose statistics of 2 patients previously treated on the CyberKnife were compared to 7-field IMRT plans. A process of acquiring images to observe intrafraction prostate motion was achieved in an average time of about 1 minute and 40 seconds, and IMRT beam delivery takes about 40 seconds per field. A complete 7-field IMRT plan can therefore be imaged and delivered in 10 to 17 minutes. The Varian Trilogy Accelerator with on-board imaging and IMRT is well suited for image-guided hypofractionated prostate treatments. During this study, we have also uncovered opportunities for improvement of the on-board imaging hardware/software implementation that would further enhance performance in this regard.