3.0 T cardiovascular magnetic resonance in patients treated with coronary stenting for myocardial infarction: evaluation of short term safety and image quality

Int J Cardiovasc Imaging. 2008 Mar;24(3):283-91. doi: 10.1007/s10554-007-9264-2. Epub 2007 Sep 6.

Abstract

Purpose: To evaluate safety and image quality of cardiovascular magnetic resonance (CMR) at 3.0 T in patients with coronary stents after myocardial infarction (MI), in comparison to the clinical standard at 1.5 T.

Methods: Twenty-five patients (21 men; 55 +/- 9 years) with first MI treated with primary stenting, underwent 18 scans at 3.0 T and 18 scans at 1.5 T. Twenty-four scans were performed 4 +/- 2 days and 12 scans 125 +/- 23 days after MI. Cine (steady-state free precession) and late gadolinium-enhanced (LGE, segmented inversion-recovery gradient echo) images were acquired. Patient safety and image artifacts were evaluated, and in 16 patients stent position was assessed during repeat catheterization. Additionally, image quality was scored from 1 (poor quality) to 4 (excellent quality).

Results: There were no clinical events within 30 days of CMR at 3.0 T or 1.5 T, and no stent migration occurred. At 3.0 T, image quality of cine studies was clinically useful in all, but not sufficient for quantitative analysis in 44% of the scans, due to stent (6/18 scans), flow (7/18 scans) and/or dark band artifacts (8/18 scans). Image quality of LGE images at 3.0 T was not sufficient for quantitative analysis in 53%, and not clinically useful in 12%. At 1.5 T, all cine and LGE images were quantitatively analyzable.

Conclusion: 3.0 T is safe in the acute and chronic phase after MI treated with primary stenting. Although cine imaging at 3.0 T is suitable for clinical use, quantitative analysis and LGE imaging is less reliable than at 1.5 T. Further optimization of pulse sequences at 3.0 T is essential.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts
  • Contrast Media
  • Female
  • Humans
  • Magnetic Resonance Imaging, Cine*
  • Male
  • Meglumine
  • Middle Aged
  • Myocardial Infarction / pathology
  • Myocardial Infarction / therapy*
  • Organometallic Compounds
  • Statistics, Nonparametric
  • Stents*
  • Treatment Outcome

Substances

  • Contrast Media
  • Organometallic Compounds
  • Meglumine
  • gadoterate meglumine