Extending the dynamic range of nuclear pulse spectrometers

Rev Sci Instrum. 2008 Mar;79(3):036105. doi: 10.1063/1.2894305.

Abstract

Using an innovative time-varying front-end electronics in conjunction with a bulky coaxial high-purity germanium detector, we were able to extend the range of the radionuclide spectra well beyond the analog to digital converter (ADC) saturation point. The electronics automatically conditions the signal for digital-filtering optimization if it is in the ADC voltage range and for time-over-threshold analysis if it exceeds the ADC range. A high spectroscopic resolution has been achieved in both operation ranges. An unprecedented wide energy range from 5 keV to 150 MeV of equivalent energy, or 90 dB, has been obtained using a single acquisition chain, while maintaining a high-energy resolution in the whole spectrum. For example, with an ADC range of 3 MeV a resolution has been obtained of 1.3/2.2 keV full width at half maximum on the 122/1332 keV gamma-ray lines of 57Co and 60Co, and of <0.4% in the time-over-threshold region, or for energy deposits beyond 3 MeV.