Calcified plaque: measurement of area at thin-section flat-panel CT and 64-section multidetector CT and comparison with histopathologic findings

Radiology. 2008 Oct;249(1):301-6. doi: 10.1148/radiol.2483072003. Epub 2008 Aug 18.

Abstract

The purpose of this study was to assess the blooming artifacts in ex vivo coronary arteries at multidetector computed tomography (CT) and flat-panel-volume CT by comparing measured areas of calcified plaque with respect to the reference standard of histopathologic findings. Three ex vivo hearts were scanned with multidetector CT and flat-panel-volume CT after institutional review board approval. The area of calcified plaque was measured at histopathologic examination, multidetector CT, and flat-panel-volume CT. The plaque area was overestimated at multidetector CT by 400% (4.61/1.15) on average, and the predicted difference between the measurements was significant (3.46 mm(2), P = .018). The average overestimation of plaque area at flat-panel-volume CT was twofold (214% [2.18/1.02]), and the predicted difference was smaller (1.16 mm(2), P = .08). The extent of the blooming artifact in visualizing calcified coronary plaque is reduced by using flat-panel-volume CT.

Publication types

  • Comparative Study

MeSH terms

  • Calcinosis / diagnostic imaging
  • Calcinosis / pathology*
  • Coronary Artery Disease / pathology
  • Coronary Vessels / pathology*
  • Heart / diagnostic imaging
  • Humans
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / methods*