Evaluation of the Cavidi ExaVir Load assay (version 3) for plasma human immunodeficiency virus type 1 load monitoring

J Clin Microbiol. 2009 Sep;47(9):3011-3. doi: 10.1128/JCM.00805-09. Epub 2009 Jul 15.

Abstract

We evaluated the new low-cost ExaVir Load (version 3) reverse transcriptase viral load assay against the Roche Cobas Amplicor assay. Results for samples tested using the reverse transcriptase assay correlated well with those obtained with the Roche assay (r = 0.85; n = 202). The version 3 reverse transcriptase assay shows improved sensitivity compared to the previous version.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Adult
  • HIV Infections / virology*
  • HIV-1 / isolation & purification*
  • Humans
  • Plasma / virology*
  • Reverse Transcriptase Polymerase Chain Reaction / methods*
  • Sensitivity and Specificity
  • Viral Load / methods*
  • Young Adult