Purpose: To perform a qualitative and quantitative comparison of different treatment planning methods used in different centres for MRI-based brachytherapy (BT) of cervical cancer.
Materials and methods: Two representative patients with advanced cervical cancer (1 "limited volume case"; 1 "extensive volume case") were planned for brachytherapy (BT) with a tandem-ring applicator by six different centres. During a workshop all centres produced an institutional standard plan and an MRI-based adaptive treatment plan for each case. Optimisation was based on the fractionation schedule (HDR, PDR) and method according to the institutional protocol.
Results: The loading pattern, dwell times, shape of the point A isodose varied considerably between institutional standard plans, as did dose-volume parameters for high risk CTV (HR-CTV) and also for the D(2cc) for OAR, violating the dose-volume constraints in many situations. During optimisation, the centres stayed as close as possible to the standard loading pattern and dwell times. The dose distributions and dose-volume parameters between the plans from the different centres became much more comparable after optimisation. The prescribed dose to the HR-CTV could be achieved in the limited volume case by all centres, in the extensive case only if additional needles were applied.
Conclusion: Treatment planning for gynaecologic brachytherapy based on different traditions shows less variation in regard to target coverage and OAR dose, when 3D image-based optimisation is performed with a uniform prescription protocol.