The imaging performance of a LaBr3-based PET scanner

Phys Med Biol. 2010 Jan 7;55(1):45-64. doi: 10.1088/0031-9155/55/1/004.

Abstract

A prototype time-of-flight (TOF) PET scanner based on cerium-doped lanthanum bromide [LaBr(3) (5% Ce)] has been developed. LaBr(3) has a high light output, excellent energy resolution and fast timing properties that have been predicted to lead to good image quality. Intrinsic performance measurements of spatial resolution, sensitivity and scatter fraction demonstrate good conventional PET performance; the results agree with previous simulation studies. Phantom measurements show the excellent image quality achievable with the prototype system. Phantom measurements and corresponding simulations show a faster and more uniform convergence rate, as well as more uniform quantification, for TOF reconstruction of the data, which have 375 ps intrinsic timing resolution, compared to non-TOF images. Measurements and simulations of a hot and cold sphere phantom show that the 7% energy resolution helps to mitigate residual errors in the scatter estimate because a high energy threshold (>480 keV) can be used to restrict the amount of scatter accepted without a loss of true events. Preliminary results with incorporation of a model of detector blurring in the iterative reconstruction algorithm not only show improved contrast recovery but also point out the importance of an accurate resolution model of the tails of LaBr(3)'s point spread function. The LaBr(3) TOF-PET scanner demonstrated the impact of superior timing and energy resolutions on image quality.

Publication types

  • Evaluation Study
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Bromides / chemistry
  • Cerium / chemistry
  • Cold Temperature
  • Computer Simulation
  • Electronics, Medical / instrumentation
  • Electronics, Medical / methods
  • Hot Temperature
  • Lanthanum / chemistry
  • Models, Theoretical
  • Phantoms, Imaging
  • Positron-Emission Tomography / instrumentation*
  • Positron-Emission Tomography / methods*
  • Scattering, Radiation
  • Sensitivity and Specificity
  • Signal Processing, Computer-Assisted / instrumentation
  • Time Factors

Substances

  • Bromides
  • lanthanum bromide
  • Cerium
  • Lanthanum