Reduction in defibrillator shocks with an implantable device combining antitachycardia pacing and shock therapy

J Am Coll Cardiol. 1991 Jul;18(1):145-51. doi: 10.1016/s0735-1097(10)80232-3.

Abstract

Implantable defibrillators reduce the risk of sudden death in patients with malignant ventricular arrhythmias, but significant restriction in quality of life can occur as a result of frequent device activation. To determine if a device that provides both antitachycardia pacing and shock therapy can safely reduce the frequency of shocks after implantation, 46 consecutive patients undergoing initial implantation of a defibrillator were studied. In all patients, the implanted device provided antitachycardia pacing and shock therapy. Detected tachycardia characteristics and the results of therapy were stored in the device's memory. There were 42 men and 4 women, aged 26 to 71 years (mean 58.7 +/- 13.5). Left ventricular ejection fraction ranged from 13% to 67% (mean 32.2 +/- 13.4%) and 31 patients had experienced one or more episodes of cardiac arrest. Induced arrhythmias included sustained monomorphic ventricular tachycardia in 38 patients, nonsustained polymorphic ventricular tachycardia in 2 and ventricular fibrillation in 4. Over a total follow-up period of 255 patient-months (range 1 to 13, mean 6.1), 25 patients experienced spontaneous arrhythmic events. In 22 patients, 909 episodes of tachycardia were treated by antitachycardia pacing, which was successful on 840 occasions (92.4%). Acceleration of ventricular tachycardia by pacing therapy was estimated to have occurred 39 times. Syncope occurred once during pacing-induced acceleration of ventricular tachycardia. Forty-four episodes of tachycardia in seven patients were treated directly by shocks because of short tachycardia cycle length; 88% of all detected tachycardias were treated without the need for shocks. Four patients died from cardiorespiratory failure and one patient died suddenly without any detected tachyarrhythmia.(ABSTRACT TRUNCATED AT 250 WORDS)

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Anti-Arrhythmia Agents / therapeutic use
  • Cardiac Pacing, Artificial / methods
  • Electric Countershock / instrumentation*
  • Electrocardiography, Ambulatory
  • Equipment Design
  • Evaluation Studies as Topic
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial*
  • Prostheses and Implants*
  • Tachycardia / mortality
  • Tachycardia / therapy*

Substances

  • Anti-Arrhythmia Agents