The direct amperometric determination of tert-butylhydroquinone (TBHQ) in biodiesel at an unmodified glassy carbon electrode is reported. A biodiesel aliquot was added into an electrochemical cell containing a 75% (v/v) ethanol-water solution under stirring (with final concentration of 50 mmol L(-1) HClO(4)). The amperometric method involved the continuous application of three sequential potential pulses to the working electrode (700 mV for 300 ms, 0 mV for 100 ms and -50 mV for 1s). TBHQ was continuously monitored at the first (direct oxidation) and optionally at the second (reduction) potential pulse while the third potential pulse was applied for cleaning of the electrode surface. For comparison, the samples were also analyzed by high-performance liquid-chromatography and a good agreement between the results was verified. Recovery values for spiked samples were between 90% and 95% and the reproducibility of the proposed method was around 5% (n=5). The proposed method can be easily adapted for on-site analysis.
Copyright (c) 2010 Elsevier B.V. All rights reserved.