Development of a short duration backlit pinhole for radiography on the National Ignition Facility

Rev Sci Instrum. 2010 Oct;81(10):10E536. doi: 10.1063/1.3496984.

Abstract

Experiments on the National Ignition Facility (NIF) will require bright, short duration, near-monochromatic x-ray backlighters for radiographic diagnosis of many high-energy density systems. This paper details a vanadium pinhole backlighter producing (1.8±0.5)×10(15) x-ray photons into 4π sr near the vanadium He-like characteristic x-ray energy of 5.18 keV. The x-ray yield was quantified from a set of Ross filters imaged to a calibrated image plate, with the Dante diagnostic used to confirm the quasimonochromatic nature of the spectrum produced. Additionally, an x-ray film image shows a source-limited image resolution of 26 μm from a 20 μm diameter pinhole.