Temporal stability of defibrillation thresholds with cardiac resynchronization therapy

Heart Rhythm. 2011 Jul;8(7):1008-13. doi: 10.1016/j.hrthm.2011.02.006. Epub 2011 Feb 9.

Abstract

Background: Defibrillation thresholds (DFTs) are typically stable over time among patients with implantable cardioverter-defibrillators (ICDs). However, the impact of cardiac resynchronization therapy (CRT) on DFTs has not been studied systematically.

Objective: This study prospectively evaluated the effect of CRT and left ventricular (LV) chamber reverse remodeling on DFTs.

Methods: This prospective, multicenter study evaluated 54 cardiac resynchronization therapy defibrillator (CRT-D) patients. Echocardiography and DFTs were performed both at implantation and at 6 months after implantation. All patients received dual-coil leads and a CRT-D pulse generator. DFTs were measured using a binary search method and tuned biphasic waveforms, where the shock pulse widths were determined by the measured shock impedance. Echocardiograms were analyzed by an independent core laboratory with a responder defined as a decrease of left ventricular end systolic volume >15%.

Results: The study cohort was 74% male, with a mean age of 68.7 ± 10.9 years. The baseline ejection fraction was 0.245 ± 0.076, and the mean New York Heart Association class was 2.9 ± 0.4. In CRT responders (n = 32) the mean DFT was 415.6 ± 108.1 V at implantation vs. 415.6 ± 124.7 V at 6 months (P = .9), and in nonresponders (n = 19) the mean DFT was 452.6 ± 102 V at implantation vs. 447.4 ± 112.4 V at 6 months (P = .8). There was no significant change in DFT peak voltage, delivered energy, or shock impedance over time.

Conclusion: DFTs were unchanged at 6 months in CRT patients with or without LV chamber reverse remodeling.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy / methods
  • Defibrillators, Implantable*
  • Echocardiography
  • Female
  • Follow-Up Studies
  • Heart Ventricles / physiopathology*
  • Humans
  • Male
  • Prospective Studies
  • Ventricular Fibrillation / diagnostic imaging
  • Ventricular Fibrillation / physiopathology
  • Ventricular Fibrillation / therapy*
  • Ventricular Remodeling / physiology*