Use of the Outreach Distal Access Catheter for microcatheter stabilization during intracranial arteriovenous malformation embolization

AJNR Am J Neuroradiol. 2012 Oct;33(9):E117-9. doi: 10.3174/ajnr.A2547. Epub 2011 Jul 14.

Abstract

The Outreach DAC is an intermediate-sized catheter designed for use with the Merci clot retriever in acute stroke. We investigated its utility as an adjunctive device during AVM pedicle embolization. In the authors' opinion, the DAC provided additional guide-catheter and microcatheter support, improved selective angiographic visualization of AVM angioarchitecture, aided microcatheter removal from its embedded position in the AVM Onyx cast, and enhanced local microcatheter control and safety, compared with embolization with the guide and microcatheter alone.

Publication types

  • Case Reports
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Drug Combinations
  • Embolization, Therapeutic / instrumentation*
  • Embolization, Therapeutic / methods
  • Equipment Design
  • Equipment Failure Analysis
  • Hemostatics / administration & dosage
  • Humans
  • Intracranial Arteriovenous Malformations / therapy*
  • Male
  • Miniaturization
  • Polyvinyls / administration & dosage*
  • Tantalum / administration & dosage*
  • Treatment Outcome
  • Vascular Access Devices*
  • Young Adult

Substances

  • Drug Combinations
  • Hemostatics
  • Onyx 18
  • Polyvinyls
  • Tantalum