Background: The currently available mechanical devices fail to achieve recanalization in as many as 20-40% of proximal arterial occlusion strokes.
Objective: The preclinical evaluation of the safety and efficacy of a novel thrombectomy device designed to achieve immediate flow restoration by quickly removing clot is reported.
Methods: Four confirmatory animal studies were performed with the Trevo device (Concentric Medical Inc, Mountain View, California, USA) in the swine (n=2) and canine (n=1) models of arterial thrombo-occlusive disease employing autologous thrombin generated thrombi. The angiographic response and the degree of device-clot incorporation were evaluated. High resolution flat panel three-dimensional CT was performed to further define the in vivo device-thrombus-vessel interaction. Finally, samples of three swine vessels treated with six passes of the device were explanted for histopathological analysis.
Results: A total of 16 clots of variable hardness and consistency were implanted in a variety of vascular settings, including the swine internal maxillary, lingual and forelimb arteries as well as the canine external carotid and vertebral arteries. Thrombolysis in Myocardial Infarction (TIMI) 2-3 reperfusion was achieved in all cases immediately after device deployment. All 16 clots were retrieved after one (n=15) or two (n=1) passes with the device. Histopathological analysis demonstrated severe disruption of the intima but no hemorrhage of media or adventitia.
Conclusion: The experimental data suggest that the Trevo device is highly effective at achieving immediate reperfusion of occluded arteries without causing any clinically significant disruption of vascular integrity.