Methicillin-resistant Staphylococcus aureus expressing low-level methicillin resistance may not be detected by the VITEK2® system

Diagn Microbiol Infect Dis. 2012 Feb;72(2):193-5. doi: 10.1016/j.diagmicrobio.2011.10.004. Epub 2011 Nov 21.

Abstract

Low-level methicillin-resistant Staphylococcus aureus may be difficult to detect with the VITEK® 2 system (VK2). Here, we suggest that S. aureus exhibiting VK2-oxacillin MIC of 1 or 2 mg/L and a negative cefoxitin screen should be tested for the presence of mecA or its gene product.