[Quality control of dose delivered by in vivo dosimetry: can one tolerance be used for all localizations?]

Cancer Radiother. 2012 Apr;16(2):115-22. doi: 10.1016/j.canrad.2011.10.010. Epub 2012 Mar 16.
[Article in French]

Abstract

Purpose: In vivo dosimetry measurements are accepted when the difference between measured and calculated dose is under 5%. A statistical analysis has been conducted to determine whether this tolerance matched the clinical practice for the studied localizations: pelvis, thorax, head and neck, breast.

Materials and methods: The technical characteristics of the detectors were checked before being used in clinical practice. Then an automatic statistical analysis was implemented using the 2450 in vivo dosimetry measurements obtained during 1 year.

Main results: The global average is 1.10%, the standard deviation 2.46% and the percentage of out of level measurements 4.09%. By distinguishing the localizations, the 5% tolerance appeared to be too narrow for the breast localization.

Discussion/conclusion: Several investigations were initiated to justify the modification of the tolerance for the breast localization. They highlighted an underestimation of the calculated dose when high beam angles are set: a new correction factor was defined to take account this error. A specific tolerance was also specified for the breast localization.

Publication types

  • English Abstract

MeSH terms

  • Decision Trees
  • Humans
  • Neoplasms / radiotherapy*
  • Quality Control*
  • Radiotherapy / adverse effects
  • Radiotherapy Dosage / standards*