Comparison of experimental and calculated calibration coefficients for a high sensitivity ionization chamber

Appl Radiat Isot. 2012 Sep;70(9):2232-6. doi: 10.1016/j.apradiso.2012.02.105. Epub 2012 Mar 3.

Abstract

The response of a Vacutec 70129 ionization chamber was calculated using the PENELOPE-2008 Monte Carlo code and compared to experimental data. The filling gas mixture composition and its pressure have been determined using IC simulated response adjustment to experimental results. The Monte Carlo simulation revealed a physical effect in the detector response to photons due to the presence of xenon in the chamber. A very good agreement is found between calculated and experimental calibration coefficients for 17 radionuclides.

Publication types

  • Comparative Study

MeSH terms

  • Calibration
  • Computer Simulation
  • Equipment Failure Analysis
  • Internationality
  • Models, Statistical*
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Radiometry / standards*
  • Reference Standards
  • Reference Values