Dose rates along the transverse axes of 125I model 6702, 125I model 6711 and 192Ir 0.2-mm steel sources for interstitial brachytherapy have been measured in a solid-water phantom for distances up to 10 cm using LiF thermoluminescent dosimeters (TLDs). Specific dose rate constants, the dose rates in water per unit source strength 1 cm along the perpendicular bisector of the source, are determined to be 0.90 +/- 0.03, 0.85 +/- 0.03, and 1.09 +/- 0.03 cGy h-1 U-1 for 125I model 6702, 125I model 6711 and 192Ir 0.2-mm steel sources, respectively (1 U = unit of air kerma strength = 1 microGy m2 h-1 = 1 cGy cm2 h-1). In older and obsolete units of source strength (i.e., mCi apparent), these are 1.14 +/- 0.03, 1.08 +/- 0.03, and 4.59 +/- 0.15 cGy h-1 mCi-1 (apparent). Currently accepted values of specific dose rate constant for 125I sources are up to 20% higher than our measured values which are in good agreement with the results of our Monte Carlo simulations. But for 192Ir there is good agreement between our measured value of the specific dose rate constant and currently accepted values. The radial dose function for 125I model 6702 is found to be consistently larger than that for 125I model 6711, with an increasing difference as the distance from the source increases. Our measured values for the radial dose function for 125I sources are in good agreement with the results of our Monte Carlo simulation as well as the measured values of Schell et al. [Int. J. Radiat. Oncol. Biol. Phys. 13, 795-799 (1987)] for model 6702 and Ling et al.(ABSTRACT TRUNCATED AT 250 WORDS)