Coronary CT angiography using the second-generation 320-detector row CT: assessment of image quality and radiation dose in various heart rates compared with the first-generation scanner

Int J Cardiovasc Imaging. 2013 Oct;29(7):1613-8. doi: 10.1007/s10554-013-0238-2. Epub 2013 May 17.

Abstract

To assess the image quality and radiation dose reduction in various heart rates in coronary CT angiography using the second-generation 320-detector row CT compared with the first-generation CT. Ninety-six patients were retrospectively included. The first 48 patients underwent coronary CT angiography with the first-generation 320-detector row CT, while the last 48 patients underwent with the second-generation CT. Subjective image quality was graded using a 4-point scale (4, excellent; 1, unable to evaluate). Image noise and contrast-to-noise ratio were also analyzed. Subgroup analysis was performed based on the heart rate. The mean effective dose was derived from the dose length product multiplied by a conversion coefficient for the chest (κ = 0.014 mSv × mGy(-1) × cm(-1)). The overall subjective image quality score showed no significant difference (3.66 vs 3.69, respectively, p = 0.25). The image quality score of the second-generation group tended to be higher than that of the first-generation group in the 66- to 75-bpm subgroup (3.36 vs 3.53, respectively, p = 0.07). No significant difference was observed in image noise and contrast-to-noise ratio. The overall radiation dose reduced by 24 % (3.3 vs 2.5 mSv, respectively, p = 0.03), and the reduction was substantial in patients with higher heart rate (66- to 75-bpm, 4.3 vs 2.2 mSv, respectively, p = 0.009; >75 bpm, 8.2 vs 3.7 mSv, respectively, p = 0.005). The second-generation 320-detector row CT could maintain the image quality while reducing the radiation dose in coronary CT angiography. The dose reduction was larger in patients with higher heart rate.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Coronary Angiography / instrumentation
  • Coronary Angiography / methods*
  • Coronary Artery Disease / diagnostic imaging*
  • Coronary Artery Disease / physiopathology
  • Coronary Vessels / diagnostic imaging*
  • Coronary Vessels / physiopathology
  • Equipment Design
  • Female
  • Heart Rate*
  • Humans
  • Male
  • Middle Aged
  • Multidetector Computed Tomography* / instrumentation
  • Predictive Value of Tests
  • Radiation Dosage*
  • Radiographic Image Interpretation, Computer-Assisted
  • Retrospective Studies
  • Signal-To-Noise Ratio
  • Tomography Scanners, X-Ray Computed