Background: Nondestructive testing of blood components could permit in-process quality control and reduce discards. Tubing segments, generated during red blood cell (RBC) component production, were tested to determine their suitability as a sample source for quality testing.
Study design and methods: Leukoreduced RBC components were produced from whole blood (WB) by two different methods: WB filtration and buffy coat (BC). Components and their corresponding segments were tested on Days 5 and 42 of hypothermic storage (HS) for spun hematocrit (Hct), hemoglobin (Hb) content, percentage hemolysis, hematologic indices, and adenosine triphosphate concentration to determine whether segment quality represents unit quality.
Results: Segment samples overestimated hemolysis on Days 5 and 42 of HS in both BC- and WB filtration-produced RBCs (p < 0.001 for all). Hct and Hb levels in the segments were also significantly different from the units at both time points for both production methods (p < 0.001 for all). Indeed, for all variables tested different results were obtained from segment and unit samples, and these differences were not consistent across production methods.
Conclusion: The quality of samples from tubing segments is not representative of the quality of the corresponding RBC unit. Segments are not suitable surrogates with which to assess RBC quality.