Accuracy and reproducibility of aortic annulus sizing using a dedicated three-dimensional computed tomography reconstruction tool in patients evaluated for transcatheter aortic valve replacement

EuroIntervention. 2014 Jul;10(3):339-46. doi: 10.4244/EIJV10I3A59.

Abstract

Aims: To evaluate the accuracy and reproducibility of aortic annulus sizing using a multislice computed tomography (MSCT) based aortic root reconstruction tool compared with conventional imaging among patients evaluated for transcatheter aortic valve replacement (TAVR).

Methods and results: Patients referred for TAVR underwent standard preprocedural assessment of aortic annulus parameters using MSCT, angiography and transoesophageal echocardiography (TEE). Three-dimensional (3D) reconstruction of MSCT images of the aortic root was performed using 3mensio (3mensio Medical Imaging BV, Bilthoven, The Netherlands), allowing for semi-automated delineation of the annular plane and assessment of annulus perimeter, area, maximum, minimum and virtual diameters derived from area and perimeter (aVD and pVD). A total of 177 patients were enrolled. We observed a good inter-observer variability of 3D reconstruction assessments with concordance coefficients for agreement of 0.91 (95% CI: 0.87-0.93) and 0.91 (0.88-0.94) for annulus perimeter and area assessments, respectively. 3D derived pVD and aVD correlated very closely with a concordance coefficient of 0.97 (0.96-0.98) with a mean difference of 0.5±0.3 mm (pVD-aVD). 3D derived pVD showed the best, but moderate concordance with diameters obtained from coronal MSCT (0.67, 0.56-0.75; 0.3±1.8 mm), and the lowest concordance with diameters obtained from TEE (0.42, 0.31-0.52; 1.9±1.9 mm).

Conclusions: MSCT-based 3D reconstruction of the aortic annulus using the 3mensio software enables accurate and reproducible assessment of aortic annulus dimensions.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging*
  • Aortic Valve Stenosis / diagnostic imaging*
  • Aortic Valve Stenosis / therapy*
  • Automation, Laboratory
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / methods
  • Echocardiography, Transesophageal
  • Female
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / methods
  • Heart Valve Prosthesis*
  • Humans
  • Imaging, Three-Dimensional*
  • Male
  • Multidetector Computed Tomography*
  • Observer Variation
  • Predictive Value of Tests
  • Prospective Studies
  • Prosthesis Design
  • Radiographic Image Interpretation, Computer-Assisted*
  • Reproducibility of Results
  • Retrospective Studies
  • Software