Objective: We aimed to compare a chemiluminescent immunoassay (CIA, QUANTA Flash) on BIO-FLASH with a multiplex flow immunoassay (MFI) on BioPlex 2200 for the detection of antibodies to Ro60, Ro52, and SS-B.
Methods: The study included 241 samples, from patients suffering from systemic autoimmune diseases (n = 108) as well as disease controls (n = 133). All samples were tested for anti-Ro52, anti-Ro60, and anti-SS-B (La) antibodies on QUANTA Flash (INOVA Diagnostics, San Diego, USA) and BioPlex 2200 (Bio-Rad Laboratories Inc., Hercules, USA). Discrepant samples were tested by two independent methods: BlueDot/ANA and QUANTRIX Microarray (both D-tek, Belgium).
Results: The overall qualitative agreements were 95.4% (95% confidence interval, CI 92.0-97.7%) for anti-Ro52, 98.8% (95% CI 96.4-99.7%) for anti-Ro60, and 91.7% (95% CI 87.5-94.9%) for anti-SS-B antibodies. There were 34 discrepant samples among all assays (20 anti-SS-B, 11 anti-Ro52, 3 anti-Ro60). 30/33 of retested samples (by D-tek dot blot) agreed with the QUANTA Flash results. Similar findings were obtained with QUANTRIX Microarray kit.
Conclusion: QUANTA Flash and BioPlex 2200 show good qualitative agreement. The clinical performances were similar for anti-Ro52 and anti-Ro60 autoantibodies while differences were observed for anti-SS-B (La) antibodies.