Next to good hemocompatibility performance of new medical devices, which has to be tested according to the ISO 10993-4, the detection of pyrogen-contaminated devices plays a pivotal role for safe device application. During blood contact with pyrogen-contaminated devices, intense inflammatory and hemostatic reactions are feared. The aim of our study was to investigate the influence of pyrogenic contaminations on stents according to the ISO 10993-4. The pyrogens of different origins like lipopolysaccharides (LPS), purified lipoteichoic acid (LTA) or zymosan were used. These pyrogens were dried on stents or dissolved and circulated in a Chandler-loop model for 90 min at 37°C with human blood. Before and after circulation, parameters of the hemostatic system including coagulation, platelets, complement and leukocyte activation were investigated. The complement system was activated by LPS isolated from Klebsiella pneumoniae and Pseudomonas aeruginosa and by LTA. Leukocyte activation was triggered by LPS isolated from K. pneumoniae, LTA and zymosan, whereas coagulation and platelet activation were only slightly influenced. Our data indicate that pyrogen-contaminated devices lead to an alteration in the hemostatic response when compared to depyrogenized devices. Therefore, pyrogenicity testing should be performed prior to hemocompatibility tests according to ISO 10993-4 in order to exclude hemostatic activation induced by pyrogen contaminations.
Keywords: Hemocompatibility; ISO 10993-4; Medical devices; Pyrogens; Surface contamination.
Copyright © 2014 Elsevier B.V. All rights reserved.