A continuum body force sensor designed for flexible surgical robotics devices

Annu Int Conf IEEE Eng Med Biol Soc. 2014:2014:3711-4. doi: 10.1109/EMBC.2014.6944429.

Abstract

This paper presents a novel three-axis force sensor based on optical photo interrupters and integrated with the robot arm STIFF-FLOP (STIFFness controllable Flexible and Learnable Manipulator for Surgical Operations) to measure external interacting forces and torques. The ring-shape bio-compatible sensor presented here embeds the distributed actuation and sensing system of the STIFF-FLOP manipulator and is applicable to the geometry of its structure as well to the structure of any other similar soft robotic manipulator. Design and calibration procedures of the device are introduced: experimental results allow defining a stiffness sensor matrix for real-time estimation of force and torque components and confirm the usefulness of the proposed optical sensing approach.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Equipment Design
  • Image Processing, Computer-Assisted
  • Linear Models
  • Minimally Invasive Surgical Procedures / instrumentation*
  • Minimally Invasive Surgical Procedures / methods
  • Robotics*
  • Torque