[St. Jude Medical valve replacement: clinical experience with 1,039 patients]

Nihon Geka Gakkai Zasshi. 1989 Sep;90(9):1513-6.
[Article in Japanese]

Abstract

St. Jude Medical valve replacement was performed in 1,039 patients; 320 had aortic (AVR), 543 mitral (MVR), and 176 had double valve replacement (DVR). There were 44(4.2%) early deaths. Follow-up extended in 995 patients from 10 to 130 months, with a cumulative period of 2,730 patients-years. The overall survival rates of AVR, MVR, and DVR patients at 10 years were 60.5%, 89.6%, 90.3% respectively. The linearized incidences of valve thrombosis, thromboembolism, anticoagulation-related hemorrhage, prosthetic valve endocarditis, and significant hemolysis were as follows: 0.11%/pt-yr, 1.33%/pt-yr, 0.04%/pt-yr, 0.18%/pt-yr, and 0.11%/pt-yr, respectively. There were no structural failure after 10 years follow-up. Reoperation (explant and re-replacement or suture repair) was required in 10 patients. Seven of them had periprosthetic leakage, 2 had valve thrombosis, and one underwent reoperation because of a technical error. Actuarially over 98% of patients were free of valve-related mortality at 10 years. St. Jude Medical valve is an excellent alternative for use in the surgical treatment of valvular heart disease.

Publication types

  • English Abstract

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aortic Valve / surgery
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis* / adverse effects
  • Humans
  • Male
  • Middle Aged
  • Mitral Valve / surgery
  • Reoperation