Four-year cardiac magnetic resonance (CMR) follow-up of patients treated with percutaneous pulmonary valve stent implantation

Eur Radiol. 2015 Dec;25(12):3606-13. doi: 10.1007/s00330-015-3781-5. Epub 2015 May 21.

Abstract

Objectives: To investigate follow-up after percutaneous pulmonary valve implantation (PPVI).

Methods: Forty patients with pulmonary conduit dysfunction (males/females 24/16; 21 ± 08 years; 12 tetralogy of Fallot, 11 aortic valve disease, 17 other congenital heart disease) were planned for CMR before PPVI and repeated 7 times up to 48 months. CMR prospective results regarded: pressure gradient (PG) and regurgitation fraction (RF); end-diastolic volume, end-systolic volume, and stroke volume indexed to body surface area (EDVI, ESVI, and SVI); ejection fraction (EF) of right and left ventricles (RV, LV). A Friedman test was used for comparisons.

Results: Overall, PG (31 ± 06 to 16 ± 4 mmHg), RF (16 ± 17 to 0.3 ± 1 %), RVEDVI (82 ± 38 to 58 ± 12 ml/m(2)), and RVESVI (44 ± 12 to 30 ± 13 ml/m(2)) declined (p < 0.001), RVEF (49 ± 13 to 58 ± 12 %) and RVSVI (from 38 ± 14 to 40 ± 8 ml/m(2)) increased (p < 0.001), LVEDVI (67 ± 17 to 73 ± 18 ml/m(2)) and LVSVI (37 ± 11 to 43 ± 10 ml/m(2)) increased (p = 0.034 and p < 0.001). Two patients had valve fracture at 24 and 36 months and underwent surgery. One patient had stent restenosis at 24 months and underwent percutaneous retreatment. Baseline/follow-up CMR did not predict PPVI failure.

Conclusions: CMR demonstrated restored pulmonary conduit function, reduced RV volumes and increased RV and LV function but did not predict valve fracture/restenosis.

Key points: • A CMR 4-year follow-up after PPVI showed restored pulmonary conduit function • RV volumes were significantly reduced • RV function was significantly better in terms of increased EF and SVI • LV function was significantly better in terms of increased EDVI and SVI • Baseline/follow-up CMR did not predict three cases of PPVI failure.

Keywords: Cardiac magnetic resonance; Congenital heart disease; Percutaneous pulmonary valve implantation; Pulmonary conduit; Right and left ventricles.

MeSH terms

  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Ventricles / pathology*
  • Heart Ventricles / physiopathology
  • Humans
  • Magnetic Resonance Imaging, Cine / methods*
  • Male
  • Prospective Studies
  • Pulmonary Valve / pathology*
  • Pulmonary Valve / surgery
  • Pulmonary Valve Insufficiency / diagnosis*
  • Pulmonary Valve Insufficiency / physiopathology
  • Pulmonary Valve Insufficiency / surgery
  • Stroke Volume / physiology*
  • Time Factors
  • Young Adult