Background: Testing for BRAF mutations in colorectal carcinoma (CRC) is important in the screening pathway for Lynch syndrome and is of prognostic value to guide management. This is a diagnostic accuracy study of the Idylla system, a novel and automated alternative PCR system.
Methods: 100 consecutive formalin-fixed, paraffin-embedded CRC resection cases were tested for BRAF mutations using the Idylla automated platform and compared with standard (Cobas) PCR.
Results: The sensitivity of the Idylla BRAF test was 100% and the specificity was 96%. Only one discordant Idylla positive/standard PCR negative result occurred and on Droplet Digital PCR demonstrated a mutation not identified by traditional PCR in this case.
Conclusion: This study has validated the Idylla system for BRAF testing in CRC and demonstrated a possibly greater sensitivity, in addition to cost effectiveness and shorter turnaround time, when compared with standard PCR.
Keywords: COLORECTAL CANCER; SURGICAL PATHOLOGY; diagnostic screening.
Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://www.bmj.com/company/products-services/rights-and-licensing/