Vascular complication can be minimized with a balloon-expandable, re-collapsible sheath in TAVR with a self-expanding bioprosthesis

Catheter Cardiovasc Interv. 2016 Jul;88(1):135-43. doi: 10.1002/ccd.26336. Epub 2015 Nov 25.

Abstract

Objective: This study evaluates the feasibility and safety of a balloon-expandable, re-collapsible sheath for TAVR patients, including those with small iliofemoral access (≤5.0 mm).

Background: The recommended iliofemoral diameter for the CoreValve TAVR system is ≥6.0 mm, but the lowest limit has not been determined.

Methods: Of 322 consecutive patients who underwent TAVR from 1/2014 to 8/2015 at two institutions, 64 underwent transfemoral CoreValve implantation, using an 11/19-French balloon-expandable, re-collapsible sheath, which has a 4.45 mm outer diameter (OD) on arterial entry, expands to 7.67 mm, then re-collapses upon removal. Valve sizing and vascular access were determined by computed tomography, and outcomes were assessed using the Valve Academic Research Consortium 2 (VARC-2) definitions.

Results: Thirteen of 64 patients had a minimal iliofemoral artery luminal diameter (MLD) of ≤5.0 mm (mean 4.38+/-0.59 mm, range 3.1-5.0 mm), with vessel calcification ≤90° to 360° and tortuosity <45° to >90°. At the MLD point, the sheath-to-artery ratios, based on the fully expanded 7.67 mm OD, ranged 1.53-2.47, higher than previously reported ratios that risk vascular complications. Major comorbidities included chronic renal failure, severe chronic obstructive pulmonary disease, extreme thrombocytopenia, cirrhosis, prior cardiac surgery, poor ventricular function, and frailty. All 64 patients had TAVR with IV sedation and local anesthesia, with 0% sheath malfunction, 0% vascular complications, and 0% bleeding in-hospital and at 30 days per VARC-2 definitions.

Conclusions: TAVR using a balloon-expandable, re-collapsible sheath is safe, including in small iliofemoral access ≤5.0 mm, thus considerably expanding the population suitable for transfemoral approach. © 2015 Wiley Periodicals, Inc.

Keywords: TAVR; VARC-2; solopath; transfemoral; vascular complication.

Publication types

  • Evaluation Study
  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / therapy*
  • Aortic Valve* / diagnostic imaging
  • Bioprosthesis*
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / instrumentation*
  • Catheterization, Peripheral / adverse effects
  • Catheterization, Peripheral / instrumentation*
  • Computed Tomography Angiography
  • Feasibility Studies
  • Female
  • Femoral Artery* / diagnostic imaging
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis*
  • Humans
  • Male
  • Middle Aged
  • Multidetector Computed Tomography
  • New York
  • Prosthesis Design
  • Retrospective Studies
  • Risk Factors
  • Severity of Illness Index
  • Texas
  • Treatment Outcome
  • Vascular Access Devices*
  • Vascular Diseases / etiology
  • Vascular Diseases / prevention & control*