Six-month outcomes of the XINSORB bioresorbable sirolimus-eluting scaffold in treating single de novo lesions in human coronary artery

Catheter Cardiovasc Interv. 2016 Mar:87 Suppl 1:630-7. doi: 10.1002/ccd.26404. Epub 2016 Feb 10.

Abstract

Objective: We aimed to investigate short-term outcomes of the XINSORB bioresorbable sirolimus-eluting scaffold in human coronary artery.

Background: Bioresorbable scaffolds are considered to be the fourth milestone in percutaneous coronary intervention.

Methods: Thirty patients with symptomatic ischemic coronary disease were enrolled and treated with the XINSORB scaffolds that were 3.0 × 12, 15, and 18 mm in size. The primary angiographic endpoint was late luminal loss (LLL), whereas the primary clinical endpoint was major adverse cardiac events (MACEs) at the 6 month follow-up. In a subset of 19 patients, intravascular ultrasound (IVUS) and optical coherence tomography (OCT) were performed at follow-up.

Results: The success rates of the procedure and the device were both 100%. Twenty-seven patients received angiographic follow-up. All patients were clinically assessed. Neither MACEs nor stent thrombus-related events were recorded. The percentage of diameter stenosis at follow-up was similar to that at postprocedure. In-scaffold and periscaffold LLL were 0.17 ± 0.12 and 0.13 ± 0.24 mm, respectively. No in-stent restenosis was detected. IVUS showed that the mean areas of the lumen, scaffold, and neointima at follow-up were 6.27 ± 0.69, 6.48 ± 0.70, and 0.20 ± 0.09 mm(2) , while in-device stenosis was 3.1 ± 1.3%. OCT showed that 97.9% of the struts presented a preserved box, while 2.1% had an open box after 6 months. A total of 95.9% of the struts were covered by neointima.

Conclusions: This first-in-human study demonstrates the effectiveness and safety of the XINSORB scaffold in treating single de novo coronary lesions.

Keywords: bioresorbable scaffold; coronary artery disease; percutaneous coronary intervention.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Absorbable Implants*
  • Adult
  • Cardiovascular Agents / administration & dosage*
  • Coated Materials, Biocompatible*
  • Coronary Angiography
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / therapy*
  • Coronary Stenosis / diagnostic imaging
  • Coronary Stenosis / therapy*
  • Coronary Vessels / diagnostic imaging
  • Coronary Vessels / drug effects*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Neointima
  • Percutaneous Coronary Intervention / adverse effects
  • Percutaneous Coronary Intervention / instrumentation*
  • Prospective Studies
  • Prosthesis Design
  • Sirolimus / administration & dosage*
  • Time Factors
  • Tomography, Optical Coherence
  • Treatment Outcome
  • Ultrasonography, Interventional

Substances

  • Cardiovascular Agents
  • Coated Materials, Biocompatible
  • Sirolimus