Vectorial total generalized variation for accelerated multi-channel multi-contrast MRI

Magn Reson Imaging. 2016 Oct;34(8):1161-70. doi: 10.1016/j.mri.2016.05.014. Epub 2016 Jun 2.

Abstract

Purpose: To develop and implement an efficient reconstruction technique to improve accelerated multi-channel multi-contrast MRI.

Theory and methods: The vectorial total generalized variation (TGV) operator is used as a regularizer for the sensitivity encoding (SENSE) technique to improve image quality of multi-channel multi-contrast MRI. The alternating direction method of multipliers (ADMM) is used to efficiently reconstruct the data. The performance of the proposed method (MC-TGV-SENSE) is assessed on two healthy volunteers at several acceleration factors.

Results: As demonstrated on the in vivo results, MC-TGV-SENSE had the lowest root-mean-square error (RMSE), highest structural similarity index, and best visual quality at all acceleration factors, compared to other methods under consideration. MC-TGV-SENSE yielded up to 17.3% relative RMSE reduction compared to the widely used total variation regularized SENSE. Furthermore, we observed that the reconstruction time of MC-TGV-SENSE is reduced by approximately a factor of two with comparable RMSEs by using the proposed ADMM-based algorithm as opposed to the more commonly used Chambolle-Pock primal-dual algorithm for the TGV-based reconstruction.

Conclusion: MC-TGV-SENSE is a better alternative than the existing reconstruction methods for accelerated multi-channel multi-contrast MRI. The proposed method exploits shared information among the images (MC), mitigates staircasing artifacts (TGV), and uses the encoding power of multiple receiver coils (SENSE).

Keywords: Alternating direction method of multipliers (ADMM); Compressed sensing; Multi-contrast magnetic resonance imaging; Parallel imaging; Total generalized variation.