Characterization of 3-Dimensional PET Systems for Accurate Quantification of Myocardial Blood Flow

J Nucl Med. 2017 Jan;58(1):103-109. doi: 10.2967/jnumed.116.174565. Epub 2016 Aug 18.

Abstract

Three-dimensional (3D) mode imaging is the current standard for PET/CT systems. Dynamic imaging for quantification of myocardial blood flow with short-lived tracers, such as 82Rb-chloride, requires accuracy to be maintained over a wide range of isotope activities and scanner counting rates. We proposed new performance standard measurements to characterize the dynamic range of PET systems for accurate quantitative imaging.

Methods: 82Rb or 13N-ammonia (1,100-3,000 MBq) was injected into the heart wall insert of an anthropomorphic torso phantom. A decaying isotope scan was obtained over 5 half-lives on 9 different 3D PET/CT systems and 1 3D/2-dimensional PET-only system. Dynamic images (28 × 15 s) were reconstructed using iterative algorithms with all corrections enabled. Dynamic range was defined as the maximum activity in the myocardial wall with less than 10% bias, from which corresponding dead-time, counting rates, and/or injected activity limits were established for each scanner. Scatter correction residual bias was estimated as the maximum cavity blood-to-myocardium activity ratio. Image quality was assessed via the coefficient of variation measuring nonuniformity of the left ventricular myocardium activity distribution.

Results: Maximum recommended injected activity/body weight, peak dead-time correction factor, counting rates, and residual scatter bias for accurate cardiac myocardial blood flow imaging were 3-14 MBq/kg, 1.5-4.0, 22-64 Mcps singles and 4-14 Mcps prompt coincidence counting rates, and 2%-10% on the investigated scanners. Nonuniformity of the myocardial activity distribution varied from 3% to 16%.

Conclusion: Accurate dynamic imaging is possible on the 10 3D PET systems if the maximum injected MBq/kg values are respected to limit peak dead-time losses during the bolus first-pass transit.

Keywords: 82Rb; cardiac positron emission tomography; dynamic range.

Publication types

  • Evaluation Study

MeSH terms

  • Blood Flow Velocity / physiology*
  • Coronary Circulation / physiology*
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods
  • Imaging, Three-Dimensional / methods*
  • Myocardial Perfusion Imaging / methods*
  • Phantoms, Imaging
  • Positron-Emission Tomography / methods*
  • Radiopharmaceuticals*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Radiopharmaceuticals