Objectives: Reoperative procedure for the treatment of a failed mitral bioprosthesis is associated with considerable risk. In some cases, mortality is high and might contraindicate the benefit of the procedure. The minimally invasive valve-in-valve (ViV) transcatheter mitral valve implant offers an alternative less-invasive approach, reducing morbidity and mortality. The objective of this paper was to evaluate the mitral ViV approach using the Braile Inovare prosthesis.
Methods: The transcatheter balloon-expandable Braile Inovare prosthesis was used in 12 cases. Procedures were performed in a hybrid operating room, under fluoroscopic and echocardiographic control. Through left minithoracotomy, the prostheses were implanted through the cardiac apex. Serial echocardiographic and clinical examinations were performed. Follow-up varied from 1 to 30 months.
Results: A total of 12 transapical mitral ViV procedures were performed. Patients had a mean age of 61.6 ± 9.9 years and 92% were women. Mean logistic EuroSCORE was 20.1%. Successful valve implantation was possible in all cases. In one case, a right lateral thoracotomy was performed for the removal of an embolized prosthesis. There was no operative mortality. Thirty-day mortality was 8.3%. Ejection fraction was preserved after the implant (66.7%; 64.8%; P = 0.3). The mitral gradient showed a significant reduction (11 mmHg; 6 mmHg; P < 0.001). Residual mitral regurgitation was not present. There was no left ventricular outflow tract obstruction.
Conclusions: The mitral ViV implant in a failed bioprosthesis is an effective procedure. This possibility might alter prosthesis selection in the future initial surgical prosthesis selection, favouring bioprostheses. Further large trials should explore its safety.
Keywords: Cardiac catheterization; Cardiopulmonary bypass; Mitral valve.
© The Author 2017. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.