Delayed right ventricular defibrillation lead perforation presenting as cardiac tamponade and treated surgically

Clin Case Rep. 2017 Mar 2;5(4):458-462. doi: 10.1002/ccr3.865. eCollection 2017 Apr.

Abstract

Right ventricular perforation leading to cardiac tamponade can occur during the chronic phase after cardiac device implantation. Physicians who manage the pacemaker clinic must be alert to the wide range of symptoms and signs that can accompany delayed right ventricular perforation. Surgical rather than percutaneous lead extraction may be prudent.

Keywords: Defibrillator lead; delayed perforation; pericardial effusion.

Publication types

  • Case Reports