Airway compliance measured by anatomic optical coherence tomography

Biomed Opt Express. 2017 Mar 15;8(4):2195-2209. doi: 10.1364/BOE.8.002195. eCollection 2017 Apr 1.

Abstract

Quantification of airway compliance can aid in the diagnosis and treatment of obstructive airway disorders by detecting regions vulnerable to collapse. Here we evaluate the ability of a swept-source anatomic optical coherence tomography (SSaOCT) system to quantify airway cross-sectional compliance (CC) by measuring changes in the luminal cross-sectional area (CSA) under physiologically relevant pressures of 10-40 cmH2O. The accuracy and precision of CC measurements are determined using simulations of non-uniform rotation distortion (NURD) endemic to endoscopic scanning, and experiments performed in a simplified tube phantom and ex vivo porcine tracheas. NURD simulations show that CC measurements are typically more accurate than that of the CSAs from which they are derived. Phantom measurements of CSA versus pressure exhibit high linearity (R2>0.99), validating the dynamic range of the SSaOCT system. Tracheas also exhibited high linearity (R2 = 0.98) suggestive of linear elasticity, while CC measurements were obtained with typically ± 12% standard error.

Keywords: (110.4500) Optical coherence tomography; (170.2150) Endoscopic imaging; (170.3880) Medical and biological imaging; (170.3890) Medical optics instrumentation.