Impact of improved attenuation correction featuring a bone atlas and truncation correction on PET quantification in whole-body PET/MR

Eur J Nucl Med Mol Imaging. 2018 Apr;45(4):642-653. doi: 10.1007/s00259-017-3864-4. Epub 2017 Nov 9.

Abstract

Purpose: Recent studies have shown an excellent correlation between PET/MR and PET/CT hybrid imaging in detecting lesions. However, a systematic underestimation of PET quantification in PET/MR has been observed. This is attributable to two methodological challenges of MR-based attenuation correction (AC): (1) lack of bone information, and (2) truncation of the MR-based AC maps (μmaps) along the patient arms. The aim of this study was to evaluate the impact of improved AC featuring a bone atlas and truncation correction on PET quantification in whole-body PET/MR.

Methods: The MR-based Dixon method provides four-compartment μmaps (background air, lungs, fat, soft tissue) which served as a reference for PET/MR AC in this study. A model-based bone atlas provided bone tissue as a fifth compartment, while the HUGE method provided truncation correction. The study population comprised 51 patients with oncological diseases, all of whom underwent a whole-body PET/MR examination. Each whole-body PET dataset was reconstructed four times using standard four-compartment μmaps, five-compartment μmaps, four-compartment μmaps + HUGE, and five-compartment μmaps + HUGE. The SUVmax for each lesion was measured to assess the impact of each μmap on PET quantification.

Results: All four μmaps in each patient provided robust results for reconstruction of the AC PET data. Overall, SUVmax was quantified in 99 tumours and lesions. Compared to the reference four-compartment μmap, the mean SUVmax of all 99 lesions increased by 1.4 ± 2.5% when bone was added, by 2.1 ± 3.5% when HUGE was added, and by 4.4 ± 5.7% when bone + HUGE was added. Larger quantification bias of up to 35% was found for single lesions when bone and truncation correction were added to the μmaps, depending on their individual location in the body.

Conclusion: The novel AC method, featuring a bone model and truncation correction, improved PET quantification in whole-body PET/MR imaging. Short reconstruction times, straightforward reconstruction workflow, and robust AC quality justify further routine clinical application of this method.

Keywords: Attenuation correction; Bone atlas; PET quantification; Truncation correction; Whole-body PET/MR.

MeSH terms

  • Adult
  • Aged
  • Bone and Bones / diagnostic imaging*
  • Female
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging*
  • Male
  • Middle Aged
  • Multimodal Imaging
  • Positron Emission Tomography Computed Tomography
  • Positron-Emission Tomography*
  • Whole Body Imaging*