Rationale and objectives: This study aims to evaluate the screening performance of digital breast tomosynthesis (DBT) combined with synthesized mammography (SM) vs combined with full-field digital mammography (FFDM).
Materials and methods: We retrospectively reviewed all screening studies utilizing FFDM + DBT (n = 7845) and SM + DBT (n = 14,776) between April 1, 2013, and February 15, 2016. Recall rate, biopsy rate, positive predictive value 1 (PPV1), positive predictive value 3 (PPV3), and cancer detection rate (CDR) were compared between the two groups. A generalized linear mixed model specifying the reading radiologist as the random effect and controlling for age was used to compare clinical outcomes between the two groups.
Results: The overall recall rate was significantly lower in the SM + DBT cohort compared to the FFDM + DBT cohort (7.06% vs 7.63%, P = .04). There was no difference in biopsy rate, PPV1, PPV3, or CDR between the two groups.
Conclusions: When DBT is performed for screening, the use of SM rather than acquiring an additional FFDM has no significant effect on biopsy rate, PPV1, PPV3, or CDR. We found a decrease in recall rate in the SM + DBT group, which may be related to the learning curve of interpreting DBT. These findings support the use of SM for patients undergoing screening with DBT.
Keywords: Mammography; breast neoplasms; early detection of cancer.