Purpose: To present a clinical validation of a novel technology called "3X" which allows for 3D prosthesis planning and treatment evaluation in total knee arthroplasty (TKA) using only 2D X-ray radiographs.
Materials and methods: After local institution review board approvals, 3X was evaluated on 43 cases (23 for preoperative planning and 20 for postoperative treatment evaluation). All the patients underwent CT scans according to a standard protocol. The results measured on the CT data were regarded as the ground truth. Additionally, two X-ray images were acquired for each affected leg and were used by 3X technology to derive patient-specific measurements of the leg. In total, we compared seven parameters for planning TKA and five parameters for postoperative prosthesis alignment.
Results: Our experimental results demonstrated that the mean distances between the surface models reconstructed from 2D X-rays and the associated surface models obtained from 3D CT data were smaller than 1.5 mm. The average differences for all angular parameters were smaller than [Formula: see text]. In over 78% cases 3X technology derived the same femoral component size as the CT-based ground truth and this value went down to 70% when 3X technology was used to predict the size of tibial component.
Conclusion: 3X is a technology that allows for true 3D preoperative planning and postoperative treatment evaluation based on 2D X-ray radiographs.
Keywords: 2D–3D reconstruction; Clinical validation; Planning; Total knee arthroplasty; Treatment evaluation.