Objective: The aim of the study was to compare the mechanical behavior of interference screw tibial fixation vs. screw-plus-staple tibial fixation in an animal model.
Methods: Thirty-six pieces of swine knee specimens were selected and divided into two groups: Group 1, tibial fixation with interference screw (n = 17), and Group 2, fixation with interference screw and staple (n = 19). The models were submitted to a single cycle of tension testing. The following variables were measured: graft cross-sectional area, failure point on 10 mm (F10), yield load (Fy), and stiffness.
Results: The mean values of graft cross-sectional area, F10, Fy, and stiffness did not present significant differences between the groups.
Conclusion: The addition of a second staple-type ligament fixation device, complementing the interference screw, did not increase the mechanical safety of the system.
Objetivo: Comparar o comportamento mecânico da fixação tibial com parafuso de interferência versus parafuso de interferência com agrafe, em modelo animal.
Métodos: Foram selecionadas 36 peças de joelho suíno e divididas em dois grupos: Grupo 1, fixação tibial com parafuso de interferência (n = 17) e Grupo 2, fixação com parafuso de interferência e agrafe (n = 19). Os modelos foram submetidos a teste de ciclo único de tração. Foram mensuradas as seguintes variáveis: medida da área de seção transversal do enxerto, ponto de falha nos 10 mm (F10), yield load (Fy) e rigidez.
Resultados: Os valores médios de área de seção transversal do enxerto, F10, Fy, e rigidez não apresentaram diferenças significativas entre os grupos.
Conclusão: A adição de um segundo dispositivo de fixação ligamentar tibial tipo agrafe, complementar ao parafuso de interferência, não aumentou a segurança mecânica do sistema.
Keywords: Anterior cruciate ligament; Biomechanical phenomena; Orthopedic fixation devices; Tendons; Tibia.