Long-Term Outcome of Hearing Rehabilitation With An Active Middle Ear Implant

Laryngoscope. 2019 Feb;129(2):477-481. doi: 10.1002/lary.27513. Epub 2018 Oct 4.

Abstract

Objective: To assess the audiological and long-term medical and technical follow-up outcomes of an active middle ear implant.

Methods: This was a retrospective medical chart analysis of all patients provided with an active middle ear implant in a tertiary academic medical referral center between September 1, 1998, and July 31, 2015. Main outcome measures were medical and technical complications, revisions, reimplantations, explantations, coupling approaches, mean time of use, pre- and postoperative hearing thresholds, functional hearing gain across frequencies (250-4,000 Hz), and Freiburg monosyllablic word test at 65 dB.

Results: One hundred and three patients were identified. Fifteen were implanted bilaterally (n = 118 Vibrant Soundbridge devices [MED-EL, Innsbruck, Austria]). Seventy-seven devices were implanted for sensorineural and 41 for mixed and conductive hearing loss. Patients used the implant for 6.7 years (range 0.7 months-17.9 years) on average. Ninety-one patients (77.12%) were using the device at the end of the observation period. An overall complication rate of 16.1% was observed. The revision and explantation rates were higher for devices implanted between 2004 and 2006. The device failure rate was 3.4%. Audiological evaluation showed significant hearing gains for both hearing loss patient groups.

Conclusion: This long-term follow-up reveals the reliability of the active middle ear implant in a single center. Overall complication rate and device failure rate are acceptable. The complication rate was higher during implementation of alternative coupling approaches. The audiological benefit was satisfactory in patients with all hearing loss types. The majority of implanted patients used the implant at the end of the observation period.

Level of evidence: 4 Laryngoscope, 129:477-481, 2019.

Keywords: Active middle ear implant; hearing loss; long-term outcome.

Publication types

  • Evaluation Study

MeSH terms

  • Auditory Threshold / physiology
  • Correction of Hearing Impairment / methods*
  • Female
  • Follow-Up Studies
  • Hearing / physiology
  • Hearing Loss, Conductive / physiopathology
  • Hearing Loss, Conductive / rehabilitation*
  • Hearing Loss, Mixed Conductive-Sensorineural / physiopathology
  • Hearing Loss, Mixed Conductive-Sensorineural / rehabilitation*
  • Hearing Loss, Sensorineural / physiopathology
  • Hearing Loss, Sensorineural / rehabilitation*
  • Humans
  • Male
  • Middle Aged
  • Ossicular Prosthesis / statistics & numerical data*
  • Postoperative Period
  • Retrospective Studies
  • Time Factors
  • Treatment Outcome