Initial experience of a large, self-expanding, and fully recapturable transcatheter aortic valve: The UK & Ireland Implanters' registry

Catheter Cardiovasc Interv. 2019 Mar 1;93(4):751-757. doi: 10.1002/ccd.27934. Epub 2018 Nov 5.

Abstract

Objectives: The UK & Ireland Implanters' registry is a multicenter registry which reports on real-world experience with novel transcatheter heart valves.

Background: The 34 mm Evolut R transcatheter aortic valve is a self-expanding and fully recapturable transcatheter aortic valve, designed to treat patients with a large aortic annulus.

Methods: Between January 2017 and April 2018, clinical, procedural and 30-day outcome data were prospectively collected from all patients receiving the 34 mm Evolut R valve across 17 participating centers in the United Kingdom and Ireland. The primary efficacy outcome was the Valve Academic Research Consortium-2(VARC-2)-defined endpoint of device success. The primary safety outcome was the VARC-2-defined composite endpoint of early safety at 30 days.

Results: A total of 217 patients underwent attempted implant. Mean age was 79.5 ± 8.8 years and Society of Thoracic Surgeons Predicted Risk of Mortality Score 5.2% ± 3.4%. Iliofemoral access was used in 91.2% of patients. Device success was 79.7%. Mean gradient was 7.0 ± 4.6 mmHg and effective orifice area 2.0 ± 0.6 cm2 . Paravalvular regurgitation was more than mild in 7.2%. A new permanent pacemaker was implanted in 15.7%. Early safety was demonstrated in 91.2%. At 30 days, all-cause mortality was 3.2%, stroke 3.7%, and major vascular complication 2.3%.

Conclusions: Real-world experience of the 34 mm Evolut R transcatheter aortic valve demonstrated acceptable procedural success, safety, valve function, and incidence of new permanent pacemaker implantation.

Keywords: aortic stenosis; self-expanding transcatheter aortic valve; transcatheter aortic valve replacement.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Female
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Ireland
  • Male
  • Postoperative Complications / mortality
  • Postoperative Complications / therapy
  • Prosthesis Design
  • Registries
  • Risk Assessment
  • Risk Factors
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome
  • United Kingdom