Progress on next generation gamma-ray Cherenkov detectors for the National Ignition Facility

Rev Sci Instrum. 2018 Oct;89(10):10I148. doi: 10.1063/1.5039378.

Abstract

Fusion reaction history and ablator areal density measurements for Inertial Confinement Fusion experiments at the National Ignition Facility are currently conducted using the Gamma Reaction History diagnostic (GRH_6m). Future Gas Cherenkov Detectors (GCDs) will ultimately provide ∼100x more sensitivity, reduce the effective temporal response from ∼100 to ∼10 ps, and lower the energy threshold from 2.9 to 1.8 MeV, relative to GRH_6m. The first phase toward next generation GCDs consisted of inserting the existing coaxial GCD-3 detector into a reentrant well which puts it within 4 m of the implosion. Reaction history and ablator gamma measurement results from this Phase I are discussed here. These results demonstrate viability for the follow-on Phases of (II) the use of a revolutionary new pulse-dilation photomultiplier tube to improve the effective measurement bandwidth by >10x relative to current PMT technology; and (III) the design of a NIF-specific "Super" GCD which will be informed by the assessment of the radiation background environment within the well described here.